
5 | Plumbing and Priming the NGC Chromatography Systems
102 | NGC Chromatography Systems and ChromLab Software
To plumb and prime the system pump wash lines
1. Mount a pump head wash tube holder to the right or left side of the instrument
in either the upper or lower rack slots.
2. Fill a 50 ml tube with 20% isopropanol (IPA) and insert it in the tube holder.
Tip: You can use 20% ethanol as an alternative.
3. Connect inlet tubing to the pump wash inlet port (left rear port) on
system pump A.
4. Immerse the other end of the pump A wash inlet tubing in the tube of 20% IPA.
5. Connect the pump head wash priming tubing to the outlet (right rear port) on
system pump B.
6. Locate a 1/4-28 female to female luer adaptor in the fittings kit and connect it to
the pump head wash priming tubing.
7. Connect a 1/4-28 union to a second inlet tubing. You will use the inlet tubing as
the outlet line for system pump B.
Tip: Set the inlet tubing with the attached union aside. Keep it close by, as
you will attach it to the pump head wash tubing in step 11.
Kommentare zu diesen Handbüchern