Model 2110FractionCollectorInstructionManualCatalog Numbers731-8120731-8122M7318120G.qxd 10/21/2002 2:47 PM Page i

Seite 2 - Warranty

Feature FunctionCarousel Holds the collection tubes. The tube positions arenumbered 1-80 on the flange of the carousel, andline up with the center of

Seite 3 - Table of Contents

4.2 Front Panel Functions Fig. 4.2. Front panel controls.Feature FunctionRun/Stop key The Run/Stop key is used to selectstandby or active mode. In the

Seite 4

Fig. 4.3 Rear Panel connections.4.3 Rear Panel FunctionsConnection FunctionPower entry Grounded 3-pin receptacle for themodule power cord. Also conta

Seite 5 - Section 1

Section 5Instrument Set-Up5.1 Tubing Installation1. Insert the drop former, narrow end downward, into the drop arm,pressing it firmly in the hole (se

Seite 6 - Introduction

abcFig. 5.1. Carousel installation.2110M7318120G.qxd 10/21/2002 2:48 PM Page 10

Seite 7 - Section 3

5.2 Carousel InstallationThe Model 2110 Fraction Collector fills tubes from the outside of thecarousel inward. Load as many tubes as desired into the

Seite 8 - Fig. 3.1. Voltage conversion

collection tube, and the eluant will flow through the carousel to waste outthe rear of the collector.5.4 Waste Collection PortAll spills on the top of

Seite 9 - Section 4

Drop Collection ModeThe fraction collector counts drops at flow rates up to ~5 ml/min. Dropcounting is performed by the sensor in the drop arm. If the

Seite 10

Time ModeThe fraction collector collects fractions at intervals of .01 minutes, fortimes per tube in the range of .05 to 9.99 minutes, and at.10 minut

Seite 11 - 4.2 Front Panel Functions

1. Attach the stopcock into the hole in the top of the drop arm. Positionthe stopcock so that its handle overhangs the end of the arm and is freeto tu

Seite 12 - 4.3 Rear Panel Functions

WarrantyModel _________________________________________Serial No. ______________________________________Date of Delivery______________________________

Seite 13 - Instrument Set-Up

7.2 Remote Operation with BioLogic LPUse System Cable 1 (catalog number 731-8261) to connect theBioLogic LP to the fraction collector. Connect the 9-p

Seite 14

Pin Function Signal Type6 External Start Input7 Advanced Contact 1 Output8 Advanced Contact 2 Output9 Logic GroundExternal Advance (Pin 1)The External

Seite 15 - 5.2 Carousel Installation

External StartThe external start line is used to allow another instrument to remote-ly start and stop the fraction collector’s programmed functions. A

Seite 16 - Stand Alone Operation

Section 9Cleaning and MaintenanceThe Model 2110 Fraction Collector requires very little maintenance toinsure reliable operation. Two procedures are ou

Seite 17

Fraction Collector Error CodesThere are several self-diagnostics built into the software of the Model2110 Fraction Collector. Below are the principal

Seite 18

ErrorNumber Cause RemedyE00 Hardware failure Unplug and replug to E01 mains power. If error per-E04 sists, contact your localBio-Rad representative.E0

Seite 19 - Section 7

Operating temperature + 4 to + 40 °CStorage temperature - 5 to + 71 °CLine voltage 90-132 VAC or 180-265 VAC, 47 Hz to 63 HzPower consumption 25 VA ma

Seite 20

Tools required: Small phillips-head screwdriver, small slot-head screw driver.Procedure:1. Disconnect the Model 2110 collector from the power source.2

Seite 21

Appendix COrdering InformationModel 2110 Fraction CollectorCatalog Number Product Description731-8122 Model 2110 Fraction Collector, 100/120 VAC (USAp

Seite 22 - AAcccceessssoorriieess

M7318120 Rev GLife ScienceGroupWeb site www.bio-rad.com USA (800) 4BIORAD Australia 02 9914 2800 Austria (01) 877 89 01 Belgium 09-385 55 11 Brazi

Seite 23 - Troubleshooting

Table of ContentsSection 1 Safety ... 1Section 2 Introduction ...

Seite 24

Appendix B Model 2110 Fraction Collector Procedurefor Adjusting Drop Detector Signal Level ... 22Appendix C Ordering Information ...

Seite 25 - Technical Specifications

Section 1Safety Disconnect supply before servicing. No user serviceable parts inside,refer servicing to Bio-Rad service personnel.High leakage current

Seite 26 - Signal Level

Section 2Introduction Fig. 2.1. Model 2110 Fraction Collector.The Model 2110 Fraction Collector (Figure 2.1), shown with theoptional instrument dust c

Seite 27

carousel where tube changes are initiated upon reaching a predetermineddrop count (1-999 drops) or time interval (0.05-99.9 minutes), or uponcommand f

Seite 28 - Ordering Information

4. Reinsert the fuse drawer in the power entry module, with the lockingtab to the left. The voltage indicator will read right-side-up if thedrawer is

Seite 29 - Laboratories

Section 4Physical Description and Control Features4.1 Physical DescriptionFig 4.1. Model 2110 Fraction Collector features.Feature FunctionDrop arm Di