Seite 1 - Instruction

For Technical Support, Call Your Local Bio-Rad Office, or in the US,Call 1-800-4BIORAD (1-800-424-6723)BioLogic™LPStarter KitInstructionManualCatalog

Seite 2

2. Connect the fraction collector to the BioLogic LP system using systemcable 3 as shown in Figure 3. Note: Both ends of cable 3 are identical.a. Conn

Seite 3 - Table of Contents

a. An SV-3 solenoid valve is used with the Model 2110 fraction collector.b. Plug the SV-3 valve directly into the diverter valve receptacle onthe back

Seite 4

3. Install tubing in pump as shown above.4. PharMed tubing, 1.6 mm ID (furnished with the BioLogic LP system), isrecommended for use with the starter

Seite 5 - Starter Kit Components

Below are the recommended tubing lengths when installing the instrument:• SV-3 to Waste [31 inches (78 cm)]• Pump Outlet to MV-6 [22 inches (57 cm)]•

Seite 6 - Buffer and Sample Preparation

2. Connect the cartridge between the injector valve outlet and the UVmonitor flow cell, as shown in the system plumbing diagram (Fig. 5). 3. Install 5

Seite 7 - System Setup

b. Press the soft key for CALIBRATE.c. Enter the Conductivity Cell Constant displayed on the conductivityflow cell cable. 2. In Manual mode: b. Press

Seite 8 - Connecting a Chart Recorder

c. Select “BioFrac” as your fraction collector. If your BioLogic LP system does not list the BioFrac, select 2128. The BioFrac communicates with the B

Seite 9

In Manual mode: a. Press the PUMP soft key, then press FLOW. b. Set flow rate to 1.5 ml/minute, press OK. Press START. c. Press the UV soft key, press

Seite 10 - Collector

Section 19Setting the AlarmThe alarm sounds after 3 minutes to remind you to turn the valve backto the load position. (Proper technique is to switch t

Seite 11 - Pump Setup

Fig. 7.Section 22Final Check Before Starting Method1. Press the MANUAL soft key, then pres the START soft key to start thepump.2. Make sure that “A” b

Seite 12 - System Tubing

4006073D.qxp 8/17/2009 7:45 AM Page CVR2

Seite 13 - Section 12

e. Chart Recorder: After zeroing the UV monitor, use the thumbwheelsto set both pens near the bottom of the chart recorder paper. Setpaper speed (12 c

Seite 14 - Section 14

Example LP Data View ChromatogramExample of chromatogram when running the BioLogic LP starter kit.Section 24Advanced Users: Automatic SampleInjectionI

Seite 15 - Section 16

3. Enter a pump program as described above—you may wish to recallyour earlier method and edit.Step 1. Buffer “A”, step Length 1 minute, flow rate 1.5

Seite 16 - Setting Up the UV Monitor

Catalog # Description731-8338 BioLogic LP System with BioFrac Fraction Collectorand LP Data View Software, 110/120 V, includes systemcables 3 and 15,

Seite 17 - Programming the Pump Method

CatalogNumber Product Information741-0018 BioFrac Prep-20 Preparative Rack741-0007 BioFrac Fraction Collector Fittings Kit, includes replacementfittin

Seite 18

Catalog # Description750-0268 BioLogic Rack Expansion Kit, includes 2 trays, 2 vertical bars, 16 sleeves 750-0261 BioLogic Rack Tray, includes 1 rack

Seite 19 - Section 22

Catalog # Description731-8233 Male Luer Plugs, 25731-8229 Female Luer T-Connector, 10732-8302 0.8 mm Barb T-Connector, recommended for minimaldead-vol

Seite 20 - Starting the Method

Legal NoticesTygon is a trademark of Norton Co. Windows and Windows 95, 98, ME, NT, 2000, and XP are trademarksof Microsoft Corp.234006073D.qxp 8/17/

Seite 21 - Injection

Life ScienceGroup00-0000 0000 Sig 0308Bulletin 0000 Rev A US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 4BIORAD Australi

Seite 22 - BioLogic LP Systems

Table of ContentsSection 1 Introduction...1Section 2 Starter Kit Components ...

Seite 23

4006073D.qxp 8/17/2009 7:45 AM Page ii

Seite 24 - BioLogic Rack

Section 1IntroductionThis manual guides you through the installation of your new BioLogicLP chromatography system. You will perform an actual chromato

Seite 25 - BioLogic LP System Fittings

2Section 3Buffer and Sample PreparationBuffer “A”1. Empty the contents of the bottle labeled “Buffer A” into a 500 ml graduated cylinder. 2. Add degas

Seite 26 - BioLogic LP System Tubing

Section 4System Setup1. Assemble the rack as shown in Figure 1. Be sure to install the racktray drain fitting, and provide a drain tube in case of spi

Seite 27 - Legal Notices

Fig. 2.Section 5Connecting a Chart Recorder1. If you will be recording data with a Bio-Rad 1327 chart recorder (discontinued item), set all chart reco

Seite 28 - Laboratories, Inc

b. System Cable 4 (Mini-DIN/Banana Plugs) connects betweenCOND CHART and the channel 2 banana plug receptacles on the 1327. The red banana plug connec