Notes: 1. If an out-of-range absorbance is measured in one of the blank wells, then the report will show:Blank mean *.***Std. Dev. *.***If the out-of

Seite 3 - Table of Contents

Notes: 1. If one of the positive or negative control wells has an out-of-range absorbance value, the values reported for positive and negative mean an

Seite 4 - Introduction

Format No. 1:Format No. 2:9Stored No. of Standard Sample Max. No.Plate Format Standards Replicates Replicates SamplesNo. 1 0–7 1 1 88No. 2 0–7 1 2 44N

Seite 5 - Product Description

Format No. 3:Format No. 4:10

Seite 6 - 2.3 Membrane Keypad

Format No. 5:Format No. 6:Notes:1. Blank wells that have been defined under the SET BLANKS mode are replaced by those of the templates. In order to se

Seite 7 - Operation

4. Under three different sets of conditions, an error may be made in cal-culation of the calibration curve and an error message will be printed on the

Seite 8 - Memory Back-up

4.2 Quick Guide—Reading a plate:1. Open the rear compartment and verify that the correct interference filters are properlyinstalled in the filter whee

Seite 9

4.3 Detailed OperationSet Analysis ModeIn this series of screens the user chooses single or dual wavelength measurement, assignsthe measurement and re

Seite 10

5. Filter selection:a. Use either ARROW to move the cursor under the check box next to Filt.b. Press SELECT to darken the check box next to Filt.c. Pr

Seite 11

Set Report Types ModeChoose the reports to be generated and automatically printed from within this mode.1. Press MODE until the display reads2. Press

Seite 12

WarrantyThis warranty may vary outside of the continental United States. Contact your local Bio-Rad office for the exact terms of your warranty.Bio-Ra

Seite 13 - Format No. 4:

3. Press ENTER. The display changes to readNote: If the display instead changes toit is because the parameters have been set for a Concentration repor

Seite 14

d. If a column has previously been defined as blank, its check box will be darkened. If a column is not yet defined as blank, then its check box is em

Seite 15 - RS-232 Interface

e. When the desired row is chosen, press NEXT to move the cursor to the column position. f. Again, use either ARROW to step up or down between 1 and 1

Seite 16

4. The cursor begins under the first digit of the Upper limit. Use either ARROW to change the first digit. 5. When the first digit is set, press NEXT

Seite 17 - 4.3 Detailed Operation

5. In this screen, choose the type of cutoff report desired.a. Use either ARROW to move the cursor under the appropriate check box.b. Press SELECT to

Seite 18

10. If it is not necessary to change the number of controls, skip to Step 12. To change the number of positive and negative controls:a. Use either ARR

Seite 19 - Set Blanks Mode

14. Negative control locations are defined just as those for positive controls. a. Use either ARROW to move the cursor under the check box next to Neg

Seite 20

d. First set the number of standards by using either ARROW to move the cursor under the check box next to STDs.e. Press SELECT to darken the check box

Seite 21 - ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■

q. Press NEXT to move the cursor to the second digit. Use either ARROW to set the second digit.r. Repeat step q until all four digits are set correctl

Seite 22 - Set Limits Mode

j. To set the number of replicates for each sample, use either ARROW to move the cursor under the check box next to Sample Repl.k. Press SELECT to dar

Seite 23 - Set Cutoff Mode

Table of ContentsSection 1 Introduction...11.1 Access

Seite 24

c. Press ENTER and the screen will changed. Use either ARROW to move the cursor under the check box next to Print Reports.e. Press SELECT to darken th

Seite 25

Section 5Instrument Service by the UserCaution: Electrical shock hazard! Always disconnect the instrument fromthe AC power source before opening the s

Seite 26 - Set Concentration Mode

Section 6Troubleshooting and Error MessagesThe Model 550 Microplate Reader constantly monitors several vital functions and will dis-play appropriate m

Seite 27

b.The best measurement wavelength Check references for the substrate you are filter for the substrate in use has not using, or determine the best mea

Seite 28

Section 7Specifications7.1 Instrument SpecificationsOperating panel Membrane keypad with 2 arrow and 6 function keysDisplay 2 lines x 16 characters LC

Seite 29 - Printing Reports

7.2 RS-232 Interface SpecificationsThe following specification describes the syntax and language required for computer con-trol of the reader. In addi

Seite 30

7.3 Command Language for the Model 550 Microplate ReaderSYNTAX:<device name><space><command>[ <space><command args.>] &l

Seite 31 - Section 5

8. <command>: “RPLATE” - read plate; arguments <mix>, <wp1>,<wp2><cmd args>: mix - Mixing time before reading (in second

Seite 32 - Section 6

Error Codes“0000” - No error“8071” - Invalid Command“8072” - Parameter Out of Range“8073” - Device Not In Remote Mode“8074” - Device Busy“8075” - Not

Seite 33

Life ScienceGroupWeb site www.bio-rad.com Bio-Rad Laboratories Main Office 2000 Alfred Nobel Drive, Hercules, CA 94547, Ph. (510) 741-1000, Fx. (510)

Seite 34 - Specifications

Section 1IntroductionThe Model 550 Microplate Reader is an eight-channel, vertical pathlength photometerthat measures the absorbance of the contents o

Seite 35

Section 2Product Description2.1 Contents of Shipping CartonThe shipping carton contains the following items:1. Model 550 Microplate Reader, with 415 n

Seite 36

Fig. 2. Rear View2.3 Membrane KeypadFig. 3. Keypad1. START/STOP Initiates plate reading using current analysis mode. Stops plate reading.2. PAPER FEED

Seite 37 - Absorbance Data

Section 3Instrument Set-up3.1 Initial Start-up1. Place the instrument on a clean, sturdy table or bench. It is important to keep the instru-ment in a

Seite 38 - Error Codes

In general, the on-board software is operated by:* Pressing MODE and the ARROW keys until the desired mode is displayed.* Pressing the ARROW keys to c

Seite 39 - Laboratories

The first time the instrument is turned on, or after a battery failure, the following defaultinformation will be held in the memory.1. 96 absorbance v