Bio-rad Model 550, Ultramark, and Benchmark Microplate Rea Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Bio-rad Model 550, Ultramark, and Benchmark Microplate Rea herunter. Bio-Rad Model 550, Ultramark, and Benchmark Microplate Reader Accessories User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - User Manual

410-0164 Rev B Ultramark™ Microplate Reader User Manual

Seite 2

4 2.4 Location of the Instrument The Ultramark microplate reader is a precise optical measuring instrument. The following conditions must be met to e

Seite 3 - Warranty Statement

5 3. Using the Ultramark The Ultramark is operated through the Protocol dialog boxes in Microplate Manager. See the Microplate Manager User Guide for

Seite 5 - Table of Contents

7 4. Changing Filters in the Ultramark This section describes how to change a filter or filters in the Ultramark. To enter filter removing mode, hold

Seite 6

8 Microplate Manager (see the Microplate Manager User Guide, section 2.7). Incorrect wavelength and clip information could result in damage to the Ult

Seite 7 - 1. General Information

9 5. Recovering from a Filter Error There are three filter error messages that can appear in the front panel of the Ultramark: • Filter Memory Error

Seite 9 - 2.3 Power Connection

11 6. Recovering a Dropped Filter To recover a filter that has dropped inside the instrument, perform the following steps: 1. Turn power o

Seite 10

12 3. Remove thumbscrew holding filter magazine in place. 4. Remove filter magazine. 5. Reach inside instrument, locate dropped filter, and

Seite 11 - 3. Using the Ultramark

13 7. Disinfection Procedure All parts of this instrument that can come in contact with patient sera or positive samples must be handled as hazardous

Seite 12

Ultramark Catalog Number 170-9500 Ultramark User Manual Catalog Number 170-9502 Bio-Rad Technical Services Department Open Monday–Friday, 8:00 a.m.

Seite 14

15 8. Error Codes The following error codes appear in the message window of the Ultramark. Error Message Probable Cause Recommended Solution Memory

Seite 15

16 Error Message Probable Cause Recommended Solution Xe flash The Xenon Lamp did not flash. If this occurs only a few times a year, it is no problem

Seite 16

17 Error Message Probable Cause Recommended Solution SCSI ini. SCSI chip trouble. Contact Bio-Rad. WCS R/W SCSI SCSI chip trouble. Contact Bio-Rad

Seite 18

19 9. Product Information 170-9500 Ultramark Microplate Imaging System, 110/230 V, Includes 10 Filters (260, 280, 340, 405, 415, 490, 570, 630, 655,

Seite 20

21 Order Form—Emission Filter Ultramark Microplate Reader Omega series filters (see website www.omegafilters.com for more information) Mechanical Sp

Seite 22

23 Disinfection Certificate This instrument and its inventory have: (1) never been in contact with any dangerous biological material, or (2) been disi

Seite 23

iii Warranty and Regulatory Notices Warranty Statement This warranty may vary outside of the continental United States. Contact your local Bio-Rad off

Seite 27 - Order Form—Emission Filter

v Table of Contents 1. General Information...1 1.1 Intr

Seite 29 - Disinfection Certificate

1 1. General Information 1.1 Introduction The Ultramark microplate reader provides all the flexibility you need for virtually any absorbance assay. Yo

Seite 31

3 2. Equipment Set-up and Installation 2.1 Unpacking the Ultramark Microplate Reader Before opening the box, check for any outer damage. Document an

Kommentare zu diesen Handbüchern