Figure 3 shows how the flow cell and the filters can be changed. To remove the flow cell, you will need to
remove the retaining clip from the bottom of the optics module and pull the luer connectors away from their
holders. Then loosen the thumbscrews and remove the flow cell. When inserting the replacement flow cell,
make sure the O-ring is securely placed on the flow cell. To change the filter, loosen the filter’s thumbscrew
and lift out and rotate the filter holder. The selected wavelength is indicated by the raised arrow on the bot-
tom of the case.
WARNING: To avoid exposure to UV radiation, turn off the UV lamp when changing filters and flow cells.
2.4 CONDUCTIVITY DETECTOR
The Conductivity Detector consists of a small, portable Conductivity Flow Cell and the control circuitry which
resides in the BioLogic LP Controller. This monitoring system provides a means to check equilibration con-
ditions and the formation of a salt gradient, for example, in ion-exchange chromatography and hydrophobic
interaction chromatography. This data helps in optimizing purification protocols and column cleaning proce-
dures.
Figure 4. Conductivity Flow Cell and Separate Holder
The Conductivity Flow Cell consists of the following:
Flow cell with Inlet/Outlet ports: For connection of the tubing, using luer fittings. The flow cell allows
flow in either direction.
Signal Cable: For connection to the back of the BioLogic LP Controller. Electrical power for the
Conductivity Flow Cell is drawn off the signal cable.
The flow cell can be plumbed immediately after the UV Optics Module’s flow cell or at any other point in the
flowpath. A separate holder for the Conductivity Flow Cell is mounted to the system rack. The swept vol-
ume in the cell is a nominal 2 µl.
Conductivity is measured in milli-Siemens (mS). For accurate measurement, the flow cell must be calibrat-
ed. Refer to Table 10.
Nine fixed Conductivity ranges, from 500 to 0.5 mS Full Scale (mSFS), are available. The user may choose
a “custom” range using the “Min/Max” function. Refer to Table 10.
11
Description of Components
Seitenansicht 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 105 106

Kommentare zu diesen Handbüchern

Keine Kommentare

Commell LV-66A manuals

Owner’s manuals and user’s guides for Hardware Commell LV-66A.
We providing 1 pdf manuals Commell LV-66A for download free by document types: User Manual






More products and manuals for Hardware Commell

Models Document Type
LV-669 User Manual   COMMELL LV-669 User Manual, 58 pages
LP-174 User Manual   COMMELL LP-174 User Manual, 25 pages
LV-667 User Manual   COMMELL LV-667 User Manual, 58 pages
LV-651 User Manual   COMMELL LV-651 User Manual, 51 pages
LN-D70 User Manual   COMMELL LN-D70 User Manual, 43 pages
LP-180 User Manual   COMMELL LP-180 User Manual, 49 pages
LP-173 User Manual   COMMELL LP-173 User Manual, 34 pages
LP-171 User Manual   COMMELL LP-171 User Manual, 39 pages
CES-471 User Manual   COMMELL CES-471 User Manual, 24 pages
LP-172 User Manual   COMMELL LP-172 User Manual, 46 pages
LP-170 User Manual   COMMELL LP-170 User Manual, 40 pages
QE-E70 User Manual   COMMELL QE-E70 User Manual, 32 pages
CES-470E User Manual   COMMELL CES-470E User Manual, 25 pages
MPX-574D User Manual   COMMELL MPX-574D User Manual, 7 pages
MPX-210 User Manual   COMMELL MPX-210 User Manual, 8 pages
ES-270E User Manual   COMMELL ES-270E User Manual, 36 pages
MPX-643 User Manual   COMMELL MPX-643 User Manual, 13 pages
MPX-750 User Manual   COMMELL MPX-750 User Manual, 10 pages
MPX-954 User Manual   COMMELL MPX-954 User Manual, 12 pages
MPX-885 User Manual   COMMELL MPX-885 User Manual, 17 pages