Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
16016 DIOM DIG3
current
Error: Over-current
16017 DIOM DIG4
current
Error: Over-current
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug.
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16018 DIOM PWM1
status
Error: Status
16019 DIOM PWM2
status
Error: Status