PowerPac™ Universal Power SupplyInstruction ManualCatalog Number164-5070For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800
Step Procedure Description8. Set the key chirp mode Press the CHIRP soft key to turn key chirp onor off.KEY CHIRP = OFF: No sound is produced when a
Step Procedure Description4. Enter limits for the Use the soft keys to select a non-constant non-constant parameter and the keypad to enter its valu
Step Procedure Description2. Select the programming On the Home screen, press the METHODSmode soft key to display the Method List screen.3. Create a
Step Procedure Description7. Enter non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter it
3.3.2 Edit (or View) and Run a Saved MethodThis section describes how to edit, view and run a saved method.Step Procedure Description 1. Set up the p
Step Procedure Description 6. Set the non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter
Step Procedure Description 11. View the run The "Run Completed" or "Run Terminated" information screen is displayed at the end of
Step Procedure Description 5. Set the non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter
Section 4Operation with a Personal Computer and PDA The PowerPac Universal is capable of transmitting methods to and from aPDA, and transmitting run d
Step Procedure Description2. Open a method There are three ways to open a method: a. Press New Method to create and open anew method.b. Select a met
Table of ContentsPageSafetySection 1 Introduction ...11.1 Overview...
Step Procedure Description4. Edit a step a. Set the Constant mode: voltage (V), cur-rent (A) or power (W).b. Enter the constant and non-constant
Step Procedure Description2. Receive data from the Press Get Data From PowerPac to request PowerPac Universal and receive the data file from the Pow
Section 5Maintenance and Troubleshooting5.1 MaintenanceThe PowerPac Universal requires little maintenance to assure reliable operation. To clean the
If a power failure occurs during a run the following screens will be displayed. 1. Displayed if PFd = ONMORE is displayed on the Run Completedscreen w
5.2.4 Error MessagesError Message Cause SolutionERROR STOP CODE 01 1. Electrophoresis cell 1. Make sure all electrical NOLOAD DETECTED not connec
Error Message Cause SolutionERROR STOP: CODE 09 1. Loose output 1. Verify all electrical RAPID CHANGE IN R connections leading connections.CORREC
5.3 Replacing a FuseIf the power supply is plugged into a working outlet, with the power switch inthe ON position and there is no display, lights or
Appendix ASpecificationsOutput specifications: 500 V, 2500 mA, 500 WOutput range (programmable): 10–500 V, fully adjustable in 1 V increments10–2500 m
Appendix BWarranty and Ordering InformationWarrantyThe PowerPac Universal is warranted for 1 year against defects in materialsand workmanship. If any
Life ScienceGroup 0104 Sig 1103Bulletin 3085 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA (800) 4BIORAD Australia 0
SafetyCaution/WarningPowerPac power supplies use high output voltages that are electrically isolated from earthground to minimize the risk of electric
Section 1Introduction1.1 OverviewThe PowerPac Universal power supply is designed to provide constant voltage,current or power for a wide range of ele
2The PowerPac Universal has the following features:• Constant voltage, constant current, or constant power operation with automaticcrossover• LCD scre
After unpacking the PowerPac Universal, remove the plastic film from thetranslucent green top case. The plastic film may leave a residue. If so, clean
Section 2Control featuresFig. 4. Front Panel.Key DescriptionRun/Pause Starts or pauses a run. When paused, the run parameters forthe current or subseq
Section 3Setup and Operation3.1 SET UPThis section describes how the PowerPac Universal power supply is set up andconnects to an electrophoresis cell
Step Procedure Description4. Set the Power Failure Use the PFd soft key to set the Power Failure detection mode detection mode:PFd = OFF: Turns powe