Model 200/2.0Power SupplyInstructionManualCatalog Numbers165-4761 and 165-4762For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call
8. Sponge pads can not dissipate heat as effectively as Scoth Brite®gel pads. For high power,high intensity applications, it is advisable to use gel p
Section 4Troubleshooting GuideCondition Cause Remedy1. With the power supply 1a. Power cord is not properly 1a. Check connections.on, the LED fails to
Eastern Regional Office, 85A Marcus Dr., Melville, New York 11747 • Phone (516) 756-2575 • Fax (516) 756-2594Australia, Bio-Rad Laboratories Pty Limit
Table of ContentsPageSection 1 General Information...11.1 General In
NoteTo insure best performance from the Model 200/2.0 Power Supply, become fully acquainted withthese operating instructions before use.WarrantyBio-Ra
1Section 1Introduction1.1 General InformationBio-Rad’s Model 200/2.0 microprocessor controlled power supply provides constantvoltage to dual output j
1.3 SpecificationsOutput voltage 5-200 volts DC, continuously adjustableOutput current 0.01-2.0 A, continuously adjustableOutput terminals Floating d
1.4 Diagram of Front Panel Membrane Control Pad1.5 Description of Membrane Control Pad1. LED display Displays the current limit, voltage, and timer
Section 2Operation2.1 Standard Mode OperationGeneral Application: The standard mode is used in blotting applications and in nucleicacid electrophore
Turn the Power Supply OffThere are three different ways to turn the power supply off.1. Press VOLTS. The red indicator light should be on. Press LOWER
63. Press LIMIT to view the set current limit. The red indicator light above the button shouldbe on, and the set limit will be shown on the LED displa