Trans-Blot® Turbo™ Blotting SystemInstruction Manual Catalog Number170-4155
Trans-Blot Transfer PacksTraydimensions(LxW):18.0x14.5cmStackdimensions(LxW):Miniformat(7.0x8.5cm),midiformat(13.5x8.5cm),+t
Table of ContentsBio-Rad Technical Support
Section 1: IntroductionSemi-drywesternblottingisacommontechniqueinmanyresearchanddiagnosticlaboratories.Conventionalsemi-dryblottingpr
Unpacking and Setup InstructionsTheTrans-BlotTurbostarterkit,catalog#170-4155,includesthemainunit,twocassettes(basesandlids),blotroll
Important:Thisinstrumentisintendedforlaboratoryuseonly.ThisproductconformstotheclassAstandardsforelectromagneticemissionsintended
Section 2: Equipment and Reagents OverviewTheTrans-Blot®Turbo™instrumentandreagentsaredesignedtoprovidefast,efcient,andreproduciblewest
Acceptable Gel Combinations Unacceptable Gel Combinations1 2 1 2Upper bay A 1miniformatgel 2minior1midiformatgels1miniformatgel 2minior
Section 3: Trans-Blot ® Turbo™ System Transfer SetupTheTurboprotocol,whencombinedwithTrans-Blottransferpacks,provideshighlyefcienttransfe
Mini Transfer Pack (Single Mini Gel) SetupThepackagingfortheminitransferpackhasthetwoionreservoirstacksintheleftandrightwellsofthe
Thetransferpacksarevacuumsealed.Checkthatthemembranedoesnotliftawaywiththefoilpackaginglid.Ifnecessary,holdtheedgeofthemembr
a.c.b.b.a.d.Fig.9.Assemblingtheminiformattransferpack.Fig.10.Assemblingthemidiformattransferpack.a.Laytheionreservoirstackwithth
Oncethestacksarepositionedinthecassettebase,placethecassettelidonthebase.Thelidisreversible,butensurethattheelectricalcontact
3.4 Transfer Using the Turbo ProtocolThe TurbobuttonimmediatelyaccessestheMIXEDMWprogram.Itisdesignedforefcienttransferofawidevariet
Note:AruncanbepausedbypressingthecorrespondingA:STOPorB:STOPNavigationbuttonduringtherun.Theuserhasachoicetocontinuefromth
3.5 Transfer Using Preprogrammed ProtocolsAllstepsareidenticaltothoseinSections3.3and3.3.a,exceptforSection3.3,step2,whichisdescrib
Fig.17.Userprotocolsscreen.3.6 Transfer Using a New ProtocolAssemblyoftheblottingsandwichisdescribedinsections3.1and3.2.Thissectiond
3.7 Transfer Using Traditional Semi-Dry ConsumablesTypicalprocedureutilizingtheTrans-BlotTurbosystemandconventionalsemi-drywesternblotting
3.8 Optimizing Transfer ConditionsThefollowingtechniques,aloneorincombination,willincreasetransferefciency:•Usealow-percentagegelo
Section 4: AppendicesAppendix A. MaintenanceTheTrans-Blot® Turbo™systemrequireslittlemaintenance.AfteryounishusingtheTrans-BlotTurbosyst
Bio-Rad Technical SupportFor help and technical advice, please contact the Bio-Rad Technical Support department. In the United States, the Technical S
Symptom/Problem SuggestedCorrectionsPoorornoelectrophoretictransfer;excessiveproteinremainsinthegel(asdetectedbystainingthegelafter
Swirlsormissingbands;diffusetransfersPoorcontactbetweenthemembraneandthegel.Airbubbles(orexcessbufferwhenusingtraditionalsemi-dr
Overview of the Three Protocol ModesTURBO:1miniformator2miniformatgels,or1midiformatgelin7min;1Mini-PROTEAN® TGX™
LIST:Bio-Radpreprogrammedprotocolsoruser-denedprotocols TechSupport:1-800-424-6723•www.bio-rad.com21Trans-Blot®Tur
NEW:Usercaneditprotocolparametersandname,save,andruntheprotocol22 TechSupport:1-800-424-6723•www.bio-rad.comTrans-Blot® Turbo™ Blotti
Consumables and Related ProductsTrans-Blot Accessories170-4151Trans-Blot®Turbo™Cassette,single170-4152
ii TechSupport:1-800-424-6723•www.bio-rad.com
Life ScienceGroup Sig 121210020688 Rev F US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Au
Safety Warning LabelsWarning:DonotopenorattempttorepairtheTrans-Blot® Turbo™instrumentorcassettes.Doingsowillvoidyourwarrantiesand
iv TechSupport:1-800-424-6723•www.bio-rad.com
Safety ComplianceThisinstrumenthasbeentestedandfoundtobeincompliancewithallapplicablerequirementsofthefollowingsafetyandelectromag
SpecificationsTrans-Blot® Turbo™ InstrumentDimensions:(LxWxH)26.0x21.1x20.4cmWeight:8lbwithcassettes,4.5lbwithoutcassettesInputpow
Kommentare zu diesen Handbüchern