ProteinChip
®
System Check Kit
Instruction Manual
Catalog #C70-00081
For technical support call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723)
For use with the ProteinChip SELDI system, Personal
or Enterprise Edition, with embedded system
processor (ESP) version 1.1.15 or higher
10010681A.qxp 8/17/2007 11:22 AM Page FC1
Seitenansicht 0
1 2 3 4 5 6 ... 20 21

Inhaltsverzeichnis

Seite 1 - Catalog #C70-00081

ProteinChip®System Check KitInstruction ManualCatalog #C70-00081For technical support call your local Bio-Rad office, or in the US, call 1-800-4BIORAD

Seite 2

2.3 System Check Maintenance Procedures 2.3.1 Maintenance Procedure 1 –– High Voltage (HV) Conditioning This procedure helps to decontaminate the su

Seite 3 - Table of Contents

Note: The detector calibration procedure is intended to standardizeinstrument performance over time. First-time use may alter systemresponse and is no

Seite 4 - Introduction

5. If this is the first calibration performed on the instrument, or if the lastcalibration was performed more than two weeks ago, select two previousl

Seite 5

2.4 System Check Tests2.4.1 Test 1 –– Detector Sensitivity This test uses the ProteinChip detector qualification array to measurethe signal-to-noise

Seite 6 - 1.2 Storage and Handling

6. Plot the peaks with noise set to 12,000 Da:a. Select all 8 spectra.b. Click on the data analysis icon . The Analysis Settingsdialog opens.c. Open

Seite 7 - Chapter 2

7. Plot all 8 spectra and select the IgG peak using the peak selectiontool.Note: Spots with high concentrations (spots A, C, E, and G) alternatewith t

Seite 8 - 2.2 File and Protocol Setup

2.4.2 Test 2 — Mass Drift and Resolution 5.96 kD This test uses the ProteinChip peptide standard array to measure themass drift and resolution of ins

Seite 9 - Conditioning Routine

6. Run the same protocol on partition “n” of 4 for all 8 spots. Recordthe partition number on the usage chart provided with the productpackaging.7. Ma

Seite 10

8. Plot the peaks with the filtering option turned off:a. Select all 8 spectra. (Do not select the first laser optimizationspectrum.)b. Click the anal

Seite 11

9. Plot all 8 spectra. Using the peak selection tool, select the insulin peak(5.963 kD) in all spectra.10. Export the data according to the instructio

Seite 12

NoticesNo part of this publication may be reproduced or transmitted in any formor by any means, electronic or mechanical, including photocopy, recordi

Seite 13 - 2.4 System Check Tests

Appendix AData Export ProcedureUse this procedure to export the data obtained during testing to theMicrosoft Excel spreadsheet provided with the kit (

Seite 14

Life ScienceGroup00-0000 0000 Sig 1106Bulletin 0000 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 4BIORAD Austra

Seite 15

Table of ContentsChapter 1 Introduction ...11.1 Product and Protocol Description ...

Seite 16

Chapter 1Introduction1.1 Product and Protocol DescriptionThe ProteinChip system check kit provides tools for performing twomaintenance procedures (de

Seite 17

Table 1. Summary of procedures and tests.CCoonnssuummaabbllee EEssttiimmaatteedd TTiimmee TTiittllee UUsseedd RReeqquuiirreedd DDeessccrriippttiio

Seite 18

Table 1. Summary of procedures and tests (continued).CCoonnssuummaabbllee EEssttiimmaatteedd TTiimmee TTeesstt UUsseedd RReeqquuiirreedd DDeessccr

Seite 19

Chapter 2Use of the ProteinChip System Check Kit2.1 OverviewThis chapter outlines the protocols used to perform the procedures andtests that comprise

Seite 20 - Data Export Procedure

• When running protocols, you specify how many sections (partitions)the spot is divided into for sampling. You also specify which of thesepartitions s

Seite 21 - Bio-Rad

Step 3Set Up theHigh Voltage (HV)Conditioning Routine61. Insert a blank ProteinChip array into the reader. If blank arrays are notavailable, use expir